近期发现有人为的截取我司网站产品图片的行为
并在未告知的我司的情况下,以我司认证的保养厂家身份进行转载和宣传
请勿被类似网站误导!!

Product information

[Basic Knowledge about Cryopumps 6 ]

Seven key points for maintenance

To maintain good performance of a cryopump;
Although our cryopumps are designed to provide high reliability and long-time continuous operation, it is necessary to perform maintenance works on a regular basis.
It is important to perform daily checks to prevent failure.


Seven key points
[Cryopump Assembly]

1. Cryopanel temperature
(Check temperature of cryopanels)
Below 20K: Normal
Above 20K: Abnormal
2. Shield temperature
(Check the electromotive force of CA thermocouple)
-5.5mV~-7.0mV: Normal
-5.5mV以上: Abnormal
(When ambient temperature is 20°C)
3. Cryopump sound, vibration
(hearing, bodily sensation)
Periodic: Normal
Irregular: Abnormal

 

[Compressor Unit]

4. Helium charge pressure while a compressor is in operation
(Read SUPPLY pressure gauge of a compressor unit)
Charge pressures vary depending on the models. Refer to compressor instruction manual for the appropriate values.
5. Elapsed time meter(E.T.M)
(Check the total operation time on ETM of a compressor unit)
Maintenance intervals vary depending on the models. Refer to compressor instruction manual and replace adsorbers, coldhead seals or bearings.
6. Sound or vibration of a compressor unit while in operation.
(by hearing or sensing bodily)
7. Cooling water flow rate, inlet temperature (with water-cooled types) Refer to compressor unit instruction manual.
Inquiries about Products

Please feel free to contact us about our product.
Our consultants will propose plans that fit your needs.

Send an E-mail